Chapter 3 _______________________________________________________________ Installation
VAISALA_______________________________________________________________________ 37
Maximum Cable Resistance Calculation for the
Barrier (Vaisala Order Code: 210664)
General specifications of HMT360:
Supply voltage U
in
= 24 V (12 ... 35 V)
Maximum current I
out
= 20 mA
Minimum operating voltage for
HMT360
U
min
= 12 V (15 V with serial port)
Stahl 9001/51-280-091-141 (values taken from the specifications):
Rated operating voltage U
N
= 20 ... 35 V
Transmitter supply voltage U
S
= U
N
- 9.5 V, when
U
N
23.5 V
or U
S
= 14 V, when U
N
23.5 V
Maximum load R
L
350

Table of Contents