User’s Guide ______________________________________________________________________
52 ___________________________________________________________________ M010056EN-I
Example:
>asel rh t
Ch1 RH lo : 0.00 %RH
Ch1 RH hi : 100.00 %RH
Ch2 T lo: -40.00 ’C
Ch2 T hi: 100.00 ’C
>
S Scaling Analog Outputs
Syntax: S [zz aa.a bb.b]<cr>
Turn the internal DIP switch Calibration Enabled/Disabled to position
ON before the scaling the analog outputs and return it to position OFF
after scaling.
Example:
>srh 0 100
RH lo: 0.00 %RH
RH hi: 100.00 %RH
>
Adjustment Commands
Turn the internal DIP switch Calibration Enabled/Disabled to position
ON before the adjustment and return it to position OFF after the
adjustment.
CRH Relative Humidity Adjustment
Syntax: CRH<cr>
where
zz = Quantity (RH, T, Td, x, a, Tw, ppm, aw, RS, Ts)
aa.a = Lower limit of the quantity
bb.b = Upper limit of the quantity
NOTE
Read also the calibrator, for example HMK15, manual and refer to page
19 of that manual for more detailed instructions on salt bath calibration
and adjustment.