Chapter 4 ________________________________________________________________ Operation
VAISALA_______________________________________________________________________ 55
Sets the output interval when the transmitter outputs measurement
readings. The time interval is used when the continuous output is active.
Example, the output interval is set to 10 minutes:
>intv 10 min
Output interval: 10 MIN
>
PRES Setting Ambient Pressure for Calculations
Syntax: PRES [aaaa.a]<cr>
Turn the internal DIP switch Calibration Enabled/Disabled to position
ON before setting the pressure and return it to position OFF after making
the setting.
Example:
>pres 12
Pressure : 12.0 bara
>
Example:
29.9213 inHg = 29.9213 × 33.86388 = 1013.25 hPa / mbar
n = 1 ... 255
xxx = S, MIN or H
where
aaaa.a = Pressure (bar
a
)
Table 6 Pressure Conversion Chart
FROM hPa/mbar mmHg/Torr inHg atm bar psi
hPa/
mbar
1 1.333224 33.86388 1013.25 1000 68.94757
TO mmHg/
Torr
0.7500617 1 25.40000 760 750.0617 51.71493
inHg 0.02952999 0.03937008 1 29.921 29.52999 2.036021
atm 0.00098692 0.00131597 0.033422 1 0.98692 0.068046
bar 0.001 0.001333224 0.03386388 1.01325 1 0.06894757
psi 0.01450377 0.01933678 0.4911541 14.6962 14.50377 1