User’s Guide ______________________________________________________________________
74 ___________________________________________________________________ M010056EN-I
Temperature (+ Operating Pressure
Ranges)
0507-021
Figure 27 Accuracy over Temperature Range
Table 8 Temperature Specifications
Property Description / Value
HMP361 -40 ... +60 °C
HMP363 80 °C -40 ... +80 °C
HMP363 120 °C -40 ... +120 °C
HMP364 -70 ... +180 °C, 0 ... 10 MPa
(0 ... 100 bar)
HMP365 -70 ... +180 °C
HMP367 -70 ... +180 °C
HMP368 -70 ... +180 °C, 0 ... 4 MPa
(0 ... 40 bar)
Accuracy at +20 °C ± 0.2 °C
Accuracy over temperature range See graph below
Sensor Pt 1000 RTD 1/3 Class B
IEC 751