EasyManua.ls Logo

Vaisala HUMICAP HMW90 SERIES - Measurement Configuration

Vaisala HUMICAP HMW90 SERIES
110 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
User's Guide _______________________________________________________________________
44 ___________________________________________________________________ M211399EN-F
Example (display full status):
>status
Device Name : HMW92
Copyright : Copyright Vaisala Oyj 2012
SW Name : XM90
SW Model : XM9x
SW version : 1.0.3.3728
Serial number : H1840005
Address : 0
SUB FUNCTIONS
* Serial Port (COM1) *
Mode : STOP
* Error Manager (ERR) *
Status : NORMAL
Active errors : 0
* MCI communication (MCI) *
Status : NORMAL
* Analog output 1 (AOUT1) *
Quantity : x
Input range : 0.000 ... 700.000 gr/lb
Output range : 4.000 ... 20.000 mA
Output clipping : 0.00 % (4.00 ... 20.00 mA)
Valid output range : 5.00 % (3.20 ... 20.80 mA)
Error value : 3.600 mA
Input now : 17.301 gr/lb
Output now : 4.395 mA
State : Normal
* Analog output 2 (AOUT2) *
Quantity : a
Input range : 0.000 ... 10000.210 gr/ft3
Output range : 4.000 ... 20.000 mA
Output clipping : 0.00 % (4.00 ... 20.00 mA)
Valid output range : 5.00 % (3.20 ... 20.80 mA)
Error value : 3.600 mA
Input now : 1280.323 gr/ft3
Output now : 6.048 mA
State : Normal
* Measurement module (HTM10) *
Status : NORMAL
Factory date : 20120425

Table of Contents