User's Guide _______________________________________________________________________
46 ___________________________________________________________________ M211399EN-F
Show Command List
Use the HELP command to list the currently available serial commands.
If the PASS command has not been used, only the basic serial commands
are available.
HELP<cr>
Example (shows basic serial commands, advanced commands are not
enabled here):
>help
?
CALCS
ECHO
ENV
ERRT
ERRS
FORM
HELP
INTV
PASS
R
RESET
SDELAY
SEND
SNUM
STATUS
UNIT
VERS
Measurement Settings
Set Environmental Parameters
Use the ENV command to set environmental parameters that affect the
measurement. For HMW90 series transmitters you can set the ambient
pressure value that is used for pressure compensation of calculated
parameters.
ENV [pressure]<cr>
where
pressure
=
Ambient pressure in hPa.
Example:
>env 1013.3
Pressure (hPa) : 1013.3