Set Analog Output Scaling
Use the ASEL command to select the output parameter and scaling for
analog output channels.
ASEL [channel parameter lo_value hi_value]<cr>
where
channel
=
Analog output channel, 1 or 2.
parameter
=
Parameter that is output on the channel.
Available parameters are:
RH relative humidity
T temperature
Tdf dew/frost point temperature
Td dew point temperature
Tw wetbulb temperature
h enthalpy
x mixing ratio
a absolute humidity
dTd dew/frost point depression
lo_value
=
Low limit of the scaling, in the units of the selected
parameter.
hi_value
=
High limit of the scaling in the units of the selected
parameter.
Example (set channel 1 to output dewpoint, in the range -10 ... 20 °C):
>pass 9000
>asel 1 TD -10 20
Aout 1 quantity : Td (-10.00 ... 20.00 'C)