7.2 Environmental Compensation
When making adjustments, you can enter information about the environment in which you
perf
orm the adjustment. This allows compensating for environmental factors that have an
eect on the measurement (for example, temperature, pressure, or background gases). The
selection of environmental compensations available in the Indigo calibration settings
depends on the features and configuration of the connected probe.
The environmental compensation selections are available on the Configuration tab of the
Calibr
ation menu. To use environmental compensations, first enable the compensation from
the Measurement selections, and then enter the compensation in the Compensation
setpoints fields. You can also set the power-up default compensations that remain in use
also after probe reset.
Measurement
Figure 23 Measurement Selections, CO
2
Pr
obe Example
Filtering factor Defines
how much past measurements aect the output
(measurement averaging over time). For details, see probe
documentation.
Temperature
compensation mode
Select whether the probe sensor's measurement or a manually
entered setpoint is used to set the temperature compensation.
Pressure/humidity/oxygen
compensation on/o
Enable or disable the environmental compensations.
Indigo 202 User Guide M211966EN-A
48