2.8 OPTPSU cable glands and connectors
Figure 6 Cable glands and connectors on OPTPSU power supply unit
#
Label Description Cable diameter Wrench
size
1 Ground connection.
4 … 16 mm
2
(12 … 5 AWG)
conductor
13 mm
2 AC in
(OPTPSU1)
Mains power input. 100 … 240 V AC,
50 … 60 Hz, 10 A.
9 … 16 mm
(0.35 … 0.63 in)
0.5 … 10mm
2
(24 … 6 AWG)
conductor
30 mm
DC in
(OPTPSU2)
Mains power input. 110 … 220 V DC,
6.3 … 2.5 A.
9 … 16 mm
(0.35 … 0.63 in)
0.5 … 10mm
2
(24 … 6 AWG)
conductor
30 mm
3 Spare Not reserved. 5 … 10 mm
(0.20 … 0.39 in)
20 mm
4 Spare Not reserved. 5 … 10 mm
(0.20 … 0.39 in)
20 mm
5 Relay out Relay output, max 250 V AC, 10 A. 7 … 12 mm
(0.28 … 0.47 in)
24 mm
Chapter 2 – Product overview
21