EasyManua.ls Logo

VAT 612 Series - Page 58

Default Icon
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
INSTALLATION Series
612
58/116
Edition 2018-08-20 605262EE
Start optimization with P gain set to 1.0 and I gain set to 0.0.
Set chamber pressure to SP2, wait until the pressure is stable. Set pressure to SP1. If the transition from
SP2 to SP1 results in a significant pressure over shoot or even does not stabilize at all, the P gain is too
high. If there is no over shoot and the pressure reaches SP1 asymptotically and very slow, P gain is too
low.
The optimal P gain value is found if the transition from SP2 to SP1 results in a slight pressure over shoot.
It does not matter if there is still a deviation between SP1 and actual pressure.
Example:

Table of Contents

Related product manuals