Endo Easy Effi cient
®
167
en
Endo Easy Effi cient
®
Immunity aspects at r.f.
The devices is intended for use in the electromagnetic environment specified below. The customer or
the user of the navigator should assure that it is used in such an electromagnetic environment.
Immunity test
Test level
EN 60601-1-2
Compliance
Level
Electromagnetic environment - guide
RF conducted
EN 61000-4-6
3 Veff from
150 kHz to
80 MHz
3 Veff from
150 kHz to
80 MHz
Portable and mobile RF communications
equipment should be used no closer to any part
of the device, including cables, than the
recommended separation distance calculated
from that equation applicable to the frequency
of the transmitter. Recommended separation
distance
d = 1.2
from 150 kHz to 80 MHz
d = 1.2
from 80 MHz to 800 MHz
d = 2.3
from 800 MHz to 2.5 GHz
where P is the maximum output power rating
of the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in meters (m)
RF radiated
EN 61000-4-3
3 Veff from
80 MHz to
2.5 GHz
3 Veff from
80 MHz to
2.5 GHz
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should
be less than the compliance level in each frequency range. Interference may occur in the vicinity of
equipment marked with the following symbol: