Maintenance and Diagnostics
Excessive radiation fields at the detector can permanently damage the
gauge. The device has a protection feature that safeguards against
excessive radiation fields. The procedure for activating this protection is
firmware dependent. Please consult the factory to activate this feature on
your device.
√ √
Display HART message Optional
Current loop output affected
Gauge status and gauge history
Gauge Status
Diagnostic Screens in Ohmview 2000
• To check the system’s present status, click Diagnostics, Diagnostics tab.
• For historical information, click the Diagnostic History and STDZ History tabs.
Since some conditions are self-repairing such as RAM and EEPROM corruption, these
conditions may appear in history screens, but not in diagnostic screens.
Acknowledge Diagnostic Alarms
Diagnostic alarms turn off when the problem resolves except for the following reasons:
• Source wipe due
• Shutter check due
• Standardize due
If the relay is set as a diagnostic alarm, you must acknowledge all diagnostic
alarms to reset the relay.
Diagnostic Alarm Messages and Conditions
View active alarm messages on the Ohmview 2000 menu. Use the Alarms, Diagnostic
Alarm Enable tab to select alarm messages.