60 DSGH Installation and Operation Guide
Advanced functions
Procedure 4.1: To check the equipment version, serial
numbers, and temperature coefficients
Min/Max History tab
The Min/Max History displays the minimum and maximum values for
parameters since the last min/max reset.
Table 4.4
Min/Max History tab — display values
You can reset these values so they record from the time of the reset.
1. Select Gauge Info | Gauge Info.
2. The Gauge Info tab appears.
Value Description
Sensor counts The raw uncompensated counts from the detector
Aux in min/max The auxiliary input counts (if used)
Sensor
Temperature
The internal temperature of the scintillator sensor in the gauge
Last reset The date of the last min/max reset