DSGH Installation and Operation Guide 67
Advanced functions
Temperature Test
This mode manually forces the sensor’s temperature probe output to a
specified value. This is useful for verifying the scintillator sensor
temperature compensation.
Procedure 4.9: To perform a temperature test
Selecting the transmitter’s type and location
Gauge Setup tab
1. Select Diagnostics | Test | Temperature Test.
2. Click Enter.
3. Remove the gauge from control. Enter the value of the new
temperature to force.
4. Click OK.
5. The transmitter functions in this mode until it times out (1
hour), or you click Exit and OK.