36 63138-US-200110
Mounting Facilities
You can also mount the source container on a mounting plate provided by the
customer or on an L-Profile.
Figure 4.9
Mounting Facility Supplied by Customer - Hole Pattern
1. Mounting Plate
2. L-profiles
Installation Control
Measurement of Local Dose Rate
After mounting, or as soon as the radioactive emitter is mounted in the source
holder, you must measure the local dose rate in the area of the source holder and
the detector.
Depending on the respective installation, radiation can also leak
out
of the beam exit channel due to
scattering. Make sure stray
radiation is
shielded off with additional lead or steel sheets. Also,
make certain all control and off-limit areas are not accessible and
marked with warning signs.
!
CAUTION