Sensor element is the CERTEC
®
measuring cell with robust,
dependent on the process tting also front-ush, abrasion-resistant
ceramic diaphragm. The process pressure causes a capacitance
change in the measuring cell via the ceramic diaphragm. This change
is converted into an appropriate output signal and outputted as
measured value.
The CERTEC
®
measuring cell is also equipped with a temperature
sensor. The temperature value can be displayed via the indicating and
adjustment module as well as processed via the signal output (with
digital versions).
As a standard feature, the CERTEC
®
measuring cell is equipped with
a lateral, recessed seal.
Instruments with double seal have an additional front seal.
Instruments with hygienic tting are equipped with a gap-free form
seal.
4 20 mA/HART two-wire electronics for voltage supply and
measured value transmission on the same cable.
The supply voltage range can dier depending on the instrument
version.
The data for power supply are spec ied in chapter "Technical data".
The optional background lighting of the indicating and adjustment
module is powered by the sensor. A certain level of operating voltage
is required for this. You can nd the exact voltage specications in
chapter "Technical data".
The optional heating requires its own operating voltage. You can nd
details in the supplementary instructions manual "Heating for indicat-
ing and adjustment module".
This function is generally not available for approved instruments.
3.3 Operation
The instrument can be adjusted with the following adjustment media:
l With indicating and adjustment module
l with the suitable VEGA DTM in conjunction with an adjustment
software according to the FDT/DTM standard, e.g. PACTware and
PC
l with manufacturer-specic adjustment programs AMS or PDM
l With a HART handheld
3.4 Packaging, transport and storage
Your instrument was protected by packaging during transport. Its
capacity to handle normal loads during transport is assured by a test
according to DIN EN 24180.