25
6 Set up the sensor with the display and adjustment module
VEGAFLEX 81 • Modbus and Levelmaster protocol
42283-EN-140430
6 Set up the sensor with the display and
adjustment module
6.1 Adjustment volume
The display and adjustment module is only used for parameter adjust-
ment of the sensor, i.e. for adaptation to the measurement task.
The parameter adjustment of the Modbus interface is carried out via
aPCwithPACTwa re.Youcanndtheprocedureinchapter"Set up
sensorandModbusinterfacewithPACTware ".
6.2 Insert display and adjustment module
The display and adjustment module can be inserted into the sensor
andremovedagainatanytime.Yo u canchooseanyoneoffourdier-
ent positions - each displaced by 90°. It is not necessary to interrupt
the power supply.
Proceed as follows:
1. Unscrew the housing cover
2. Place the display and adjustment module in the requested posi-
tion onto the electronics and turn to the right until it snaps in
3. Screw housing cover with inspection window tightly back on
Removal is carried out in reverse order.
The display and adjustment module is powered by the sensor, an ad-
ditional connection is not necessary.
Fig. 17: Insertion of the display and adjustment module
Note:
Ifyouintendtoretrottheinstrumentwithadisplayandadjustment
module for continuous measured value indication, a higher cover with
an inspection glass is required.