54
7 Setting up sensor and Modbus interface with PACTware
VEGAFLEX 81 • Modbus and Levelmaster protocol
42283-EN-140430
Device name
Here,youcanndtheinstrumentname.Youcannotchangethisline
becausetheinstrumentnameisunmodiablysavedintheinstrument.
Serial number
Here,youcanndtheserialnumberofyourinstrument.Yo u cannot
changethislinebecausetheserialnumberisunmodiablysavedin
the instrument.
Measurement loop name
Here you can enter a suitable measurement loop name for your VE-
GAFLEX 81. You can enter a name with max. 19 characters. You can
use capital and small letters as well as numbers. The following special
characters are also possible: + - . : , ( ) / < >
Probelengthmodied?
Ifyouhavemodiedtheprobelength,thismustbeenteredinthe
selectioneld.
If you select "No", then the instrument uses automatically the
preset length of the default setting.
If you select "Yes ",thenyoucanenterinanothereldthemodied
length of the instrument.
Probe length L from seal surface
Ifyouhavemodiedthelengthoftheprobe,youcanenterinthiseld
themodiedprobelength.Keeptheselectedunitinmind.
Determine probe length automatically
If you do not know the probe length, you can have the length of the
probe determined automatically. The requirement for this is a probe
unrestricted and not covered by the medium.
Click to "Carryoutnow", to start the automatic length determination.
Type of medium
Here you can see which type of medium your instrument is suitable
for. If this function is inactive, the medium type your instrument is suit-
able for was already preset.
Information:
In special cases you can change the type of medium. This setting can
be changed under "Extendedadjustment".
Application
Inthiseldyoucanselecttheapplicationyouwanttouseyourinstru-
ment for. You have the following selection options:
Level in the vessel
Level in the bypass/standpipe
Interface in the vessel
Interface in the bypass/standpipe
Demonstration mode
Level measurement: If you select "Level", you can select the proper-
tiesofthemediuminanothereld.