8 Diagnostics and service
VEGAFLEX 81 • Modbus and Levelmaster protocol
42283-EN-140430
Code
Text mes-
sage
Cause Rectication
S600
Impermissi-
ble electronics
temperature
Temperature of the pro-
cessing electronics in the
non-speciedsection
Check ambient temperature
Isolate electronics
Use instrument with higher
temperature range
S601
Overlling
Level echo in the close
range not available
Reduce level
100 % adjustment: Increase
value
Check mounting socket
Remove possible interfering
signals in the close range
Use coaxial probe
S602
Level with-
in the search
range, com-
pensation
echo
Compensation echo super-
imposed by medium
100 % adjustment: Increase
value
S603
Impermissi-
ble operating
voltage
Operating voltage below
speciedrange
Check electrical connection
if necessary, increase
operating voltage
The following table shows the error codes and text messages in the
status message "Maintenance" and provides information on causes
as well as corrective measures.
Code
Text mes-
sage
Cause Rectication
M500
Error in the
delivery sta-
tus
With the reset to delivery
status, the data could not
be restored
Repeat reset
LoadXMLlewithsensor
data into the sensor
M501
Error in the
non-active
linearization
table
Hardware error EEPROM Exchanging the electronics
Send instrument for repair
M502
Error in the
event memory
Hardware error EEPROM Exchanging the electronics
Send instrument for repair
M503
Reliability too
low
Measurement certainty is
too low for a reliable meas-
urement
Process component or
probe contaminated or
defective
Check installation and
process conditions
Clean or exchange process
component or probe