77
10 Supplement
VEGAFLEX 81 • Modbus and Levelmaster protocol
42283-EN-140430
Materials
Ʋ Housing ABS
Ʋ Inspection window Polyester foil
Integrated clock
Date format Day.Month.Year
Time format 12 h/24 h
TimezoneExfactory CET
Rate deviation max. 10.5 min/year
Measurement electronics temerature
Resolution 1 °C (1.8 °F)
Accuracy ±1 °C (1.8 °F)
Permissible temperature range -40 … +85 °C (-40 … +185 °F)
Voltage supply
Operating voltage 8 … 30 V DC
Power consumption < 500 mW
Interpolation protection Integrated
Electrical protective measures
Protection, depending on housing version
Ʋ Plastic housing IP 66/IP 67 (NEMA 4X)
Ʋ Aluminium housing, stainless steel
housing - investment casting, stain-
less steel housing - electro-polished
IP 66/IP 68 (0.2 bar), NEMA 6P
6)
Overvoltage category III
7)
Protection class III
8)
Approvals
Instrumentswithapprovalscanhavedierenttechnicalspecicationsdependingontheversion.
For that reason the associated approval documents of these instruments have to be carefully
noted. They are part of the delivery or can be downloaded under www.vega.com, "VEGATools"
and "Instrument search" as well as under www.vega.com/downloads and "Approvals".
10.2 Basics Modbus
Bus description
The Modbus protocol is a communication protocol for the communication between instruments. It is
based on a Master/Slave or Client/Server architecture. By means of Modbus, a Master and several
Slaves can be connected. Each bus participant has an unambiguous address and can send mes-
sages to the bus. This is initiated by the Master, the addessed Slave answers. For data transfer, the
6)
The prerequisites for maintaining the protection rating are a suitable cable as well as correct mounting.
7)
IEC 61010-1
8)