16
4 Mounting
VEGAFLEX 86 • Probus PA
44232-EN-160518
1 2
Fig. 5: Installation in non-metallic vessel
1 Flange
2 Metal sheet
If possible, avoid sockets. Mount the sensor ush with the vessel top.
If this is not possible, use short sockets with small diameter.
Higher sockets or sockets with a bigger diameter can generally be
used. They can, however, increase the upper blocking distance (dead
band). Check if this is relevant for your measurement.
In such cases, always carry out a false signal suppression after instal-
lation. You can nd further information under "Setup procedure".
≤ 150 mm (5.91")
≤ 100 mm (3.94")
h
d