EasyManua.ls Logo

Vega VEGAFLEX 86 - Wiring plan, single chamber housing

Vega VEGAFLEX 86
100 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
27
5 Connecting to power supply
VEGAFLEX 86 • Probus PA
44232-EN-160518
Fig.15:Connectionsteps5and6-Doublechamberhousing
6. Insert the wire ends into the terminals according to the wiring plan
Information:
Solid cores as well as exible cores with wire end sleeves are insert-
ed directly into the terminal openings. In case of exible cores without
end sleeves, press the terminal from above with a small screwdriver,
the terminal opening is then free. When the screwdriver is released,
the terminal closes again.
You can nd further information on the max. wire cross-section under
"
Technicaldata-Electromechanicaldata"
7. Check the hold of the wires in the terminals by lightly pulling on
them
8. Connect the screen to the internal ground terminal, connect the
external ground terminal to potential equalisation
9. Tighten the compression nut of the cable entry gland. The seal
ring must completely encircle the cable
10. Reinsert the display and adjustment module, if one was installed
11. Screw the housing lid back on
The electrical connection is nished.
5.3 Wiring plan, single chamber housing
The following illustration applies to the non-Ex, Ex-ia and Ex-d-ia
version.

Table of Contents

Other manuals for Vega VEGAFLEX 86

Related product manuals