35
6 Set up with the display and adjustment module
VEGAFLEX 86 • Probus PA
44232-EN-160518
Move to the menu overview
Conrm selected menu
Edit parameter
Save value
[->] key:
Change measured value presentation
Select list entry
Select editing position
[+] key:
Change value of the parameter
[ESC] key:
Interrupt input
Jump to next higher menu
You adjust the sensor via the four keys of the display and adjustment
module. The individual menu items are shown on the LC display. You
can nd the functions of the individual keys in the previous illustration.
When the [+] and [->] keys are pressed quickly, the edited value,
or the cursor, changes one value or position at a time. If the key is
pressed longer than 1 s, the value or position changes continuously.
When the [OK] and [ESC] keys are pressed simultaneously for more
than 5 s, the display returns to the main menu. The menu language is
then switched over to "English".
Approx. 60 minutes after the last pressing of a key, an automatic reset
to measured value indication is triggered. Any values not conrmed
with [OK] will not be saved.
After switching on, the VEGAFLEX 86 carries out a short self-test
where the device software is checked.
The output signal transmits a fault signal during the switch-on phase.
The following information is displayed on the display and adjustment
module during the startup procedure:
Instrument type
Device name
Software version (SW-Ver)
Hardware version (HW-Ver)
With the [->] key you can move between three dierent indication
modes.
In the rst view, the selected measured value is displayed in large
digits.
In the second view, the selected measured value and a correspond-
ing bar graph presentation are displayed.
In the third view, the selected measured value as well as a second
selectable value, e.g. the temperature, are displayed.
Adjustment system
S
witch-on phase
Measured v
alue indica-
tion