8
3 Product description
VEGAFLEX 86 • Probus PA
44232-EN-160518
Scan the Data Matrix code on the type label of the instrument or
Enter the serial number manually in the app
This operating instructions manual applies to the following instrument
versions:
Hardware from 1.0.0
Software from 1.2.0
Only for instrument versions without SIL qualication
This electronics version can be determined via the product code on
the type label as well as on the electronics.
Standard electronics: Type FX80PA.-
The scope of delivery encompasses:
Sensor
Documentation
Quick setup guide
Test certicate measuring accuracy (optional)
Supplementary instructions "GSM/GPRSradiomodule"
(optional)
Supplementary instructions manual "Heating for display and
adjustment module" (optional)
Supplementary instructions manual "Plug connector for con-
tinuously measuring sensors" (optional)
Ex-specic "Safety instructions" (with Ex versions)
If necessary, further certicates
3.2 Principle of operation
The VEGAFLEX 86 is a level sensor with cable or rod probe for
continuous level or interface measurement, particularly suitable for
applications in high temperatures up to +450 °C (842 °F).
High frequency microwave pulses are guided along a steel cable or
a rod. Upon reaching the product surface, the microwave pulses are
reected. The running time is evaluated by the instrument and output-
ted as level.