9MeasuredvaluetransmissiontoaWEB-VVserver
RS232/Ethernet connection • VEGAMET 391/624/625, VEGASCAN 693, PLICSRADIO C62
30325-EN-130701
TheMoRoScontainsaninternalclockforcontrollingtime-dependent
eventsanddatingsystemmessages.Setthetime,selecttheproper
timezoneandactivatethetimesynchronization.
Fig. 33: Time synchronisation
9.6 Set up measured value transmission
Selectthemenuitem"Instrument settings - Event list"andpressthe
button"New".Anassistantwillnowleadyouthroughthecongura-
tion.
1. Selectintherstwindowunder"Message type"thelistentry
"WEB-VV"or"WEB-VV (encoded)".
2. InthesecondstepenterthenameortheIPaddressoftheWEB-
VVserver.Ifyou’reusingtheVEGAservers(VEGAHosting/
WEB-VVVH),entertheURL"web-vv.vega.com"here.Ifyouhave
aserverofyourown(LocalHosting/WEB-VVLH),youcanget
theservernameortheIPaddressfromyourresponsiblenetwork
administrator.
3. Inthelaststep,denewhenorhowoftenthemeasuredvalues
aretransmitted.Pleaserememberthatcostsarisewithevery
measuredvaluetransmission,dependingonthetypeofconnec-
tion.
4. Transferyoursettingstothesignalconditioninginstrument.
Aftercompletingtheconguration,youshouldstartatesttransmis-
siontotheWEB-VVserver.ThencontacttheWEB-VVadministrator.
Hewillallocatethenewinstrumentnetworkaccordinglyandrelease
themeasuredvaluesforyourinspection.
Time