10 Diagnostics and servicing
VEGAPULS 11 • Two-wire 4 … 20 mA
58350-EN-200806
Code
Text message
Cause Rectication
M505
No echo available
Sensor does not detect an echo during
operation
Antenna dirty or defective
Clean the antenna
Use a more suitable antenna/sensor
Remove possible false echoes
Optimize sensor position and orientation
M507
Error in the instrument
settings
Error during setup
Error when carrying out a reset
False signal suppression faulty
Carry out reset and repeat setup
M508
No executable Bluetooth
software
Checksum error in Bluetooth software Carry out software update
M509
Software update running
Software update running
Waituntilsoftwareupdateisnished
M510
No communication with
the main controller
Communication between main electronics
and display module disturbed
Check the connection cable to the display
Send instrument for repair
M511
Inconsistent software
conguration
A software unit requires a software update Carry out software update
10.5 Treatment of measurement errors
The tables below give typical examples of application-related meas-
urement errors.
The images in column " Error description" show the actual level as a
dashed line and the output level as a solid line.
1
2
Level
time
0
1 Real level
2 Level displayed by the sensor
Note:
If the output level is constant, the cause could also be the fault setting
of the current output to " Hold value".
If the level is too low, the reason could be a line resistance that is too
high