EasyManua.ls Logo

Vega VEGAPULS 31 - Technical Specifications and Data; Instrument Technical Data

Vega VEGAPULS 31
64 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
54
14 Supplement
VEGAPULS 31 • Two-wire 4 … 20 mA/HART
57820-EN-191014
14 Supplement
14.1 Technical data
Note for approved instruments
The technical data in the respective safety instructions which are included in delivery are valid for
approvedinstruments(e.g.withExapproval).Thesedatacandierfromthedatalistedherein,for
example regarding the process conditions or the voltage supply.
All approval documents can be downloaded from our homepage.
Materials and weights
Materials, wetted parts
Ʋ Antenna,processtting PVDF
Ʋ Process seal FKM
Materials, non-wetted parts
Ʋ Housing Plastic PBT (Polyester)
Ʋ Housing seals Silicone SI 850 R
Ʋ Cable gland PA
Ʋ Sealing, cable gland NBR, EPDM
Ʋ Blind plug, cable gland PA
Ʋ Inspection window for the indication Polycarbonate
Weight
0.7 kg (1.543 lbs)
Torques
Max. torque mounting boss
7 Nm (5.163 lbf ft)
Max. torque for NPT cable glands and
Conduit tubes
10 Nm (7.376 lbf ft)
Input variable
Measured variable The measured variable is the distance between the
antenna edge of the sensor and the product surface.
The antenna edge is also the reference plane for the
measurement.

Table of Contents

Related product manuals