EasyManua.ls Logo

Vega VEGAPULS 61 - Page 61

Vega VEGAPULS 61
100 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
61
8 Diagnosis, asset management and service
VEGAPULS 61 • Modbus and Levelmaster protocol
41717-EN-160224
Code
Text message
Cause Rectication DevSpec
Diagnosis Bits
F260
Error in the calibration
Error in the calibration
carried out in the factory
Error in the EEPROM
Exchanging the elec-
tronics
Send instrument for
repair
Bit 8
F261
Errorintheconguration
Error during setup
False signal suppression
faulty
Error when carrying out
a reset
Repeat setup
Repeat reset
Bit 9
F264
Installation/Setup error
Adjustment not within
the vessel height/meas-
uring range
Max. measuring range
of the instrument not
sucient
Check or correct instal-
lation and/or parameter
adjustment
Use an instrument with
bigger measuring range
Bit 10
F265
Measurement function dis-
turbed
Sensor no longer carries
out a measurement
Operating voltage too
low
Check operating voltage
Carry out a reset
Disconnect operating
voltagebriey
Bit 11
The following table shows the error codes and text messages in the
status message "Function check" and provides information on causes
as well as corrective measures.
Code
Text message
Cause Rectication DevSpec State in CMD
48
C700
Simulation active
A simulation is active Finish simulation
Wait for the automatic
end after 60 mins.
"Simulation Active" in
"StandardizedStatus0"
The following table shows the error codes and text messages in the
status message "Out of specication" and provides information on
causes as well as corrective measures.
Code
Text message
Cause Rectication DevSpec State in CMD
48
S600
Impermissible electronics
temperature
Temperature of the
electronics in the non-
speciedrange
Check ambient tem-
perature
Isolate electronics
Use instrument with
higher temperature
range
Bit 5 of Byte 14…24
S601
Overlling
Danger of vessel
overlling
Make sure that there is
nofurtherlling
Check level in the vessel
Bit 6 of Byte 14…24
The following table shows the error codes and text messages in the
status message "Maintenance" and provides information on causes
as well as corrective measures.
Function check
Outofspecication
Maintenance

Table of Contents

Other manuals for Vega VEGAPULS 61

Related product manuals