Technical Specications
VentecLife.com 156
Control Accuracy
NOTE: Negative (subatmospheric) pressure is not available during the expiratory phase of ventilation therapy breaths
delivered by V*Home.
NOTE: When used with active, passive, or valveless Ventec One‑Circuits, V*Home was designed for use with a humidier
or HME. All volumes and ows are expressed in BTPS unless stated otherwise.
Control Accuracy
Apnea Rate ±1 BPM or ±10% of setting, whichever is greater
Breath Rate Accuracy: ±1 breath/minute, or ±10% of setting, whichever is greater
Stability: ±10% of setting
EPAP See PEEP.
Flow ±(10% of setting or 1 L/min, whichever is greater)
Flow Cycle ±10% for active and mouthpiece circuits
±15% for passive and valveless circuits
Flow Trigger ±1 L/min when PEEP is set to 0 cmH2O
±1.5 L/min when PEEP is set to 1 to 6 cmH2O
±2 L/min when PEEP is set to 7 to 16 cmH2O
±2.2 L/min when PEEP is set from 17 to 25 cmH2O
Humidication N/A
Inspiratory Time Accuracy: ±(10% of setting + 0.1 seconds)
Stability: ±(10% of setting + 0.1 seconds)
IPAP ±(8% of setting or 2 cmH2O, whichever is greater)
Leak Compensation N/A
Nebulizer Duration ±1 minute
Patient Type N/A
Pause Time ±(10% of setting + 0.1 seconds)
PC Flow Termination N/A
PEEP/EPAP ±(10% of setting + 1 cmH2O)
Pressure Control ±(8% of setting or 2 cmH2O, whichever is greater)
Pressure Support ±(8% of setting or 2 cmH2O, whichever is greater)
Rise Time N/A
Sigh N/A