Breath Types and Therapy Modes
VentecLife.com 50
Comparable Non-Invasive Ventilation Modes
Ventilation Mode or Feature Equivalent V*Home Mode Settings Additional V*Home Controls
S/T
(Spontaneous/Timed)
Set Mode to Bi-Level
Set Breath Rate
Set Flow Trigger
Inspiratory Time
EPAP
IPAP (measured from a baseline of 0
cmH2O)
Flow Cycle
Time Cycle
Rise Time
AVAPS™, PRVS
(Average Volume Assured Pressure
Support)
(Pressure Regulated Volume
Support)
Set Mode to Vol. Targeted-PS
Set Breath Rate
Set Pres. Minimum and Pres. Adj. Rate
Set Flow Trigger
Set High Pressure Alarm 5 cmH2O
above desired maximum pressure
Inspiratory Time
PEEP
Tidal Volume
Flow Cycle
Time Cycle
Rise Time
T
(Timed)
Set Mode to AC-Pressure
Set Breath Rate
Set Flow Trigger to Off
Set Pressure Control Flow Termination
to Off
Inspiratory Time
PEEP
Pressure Control (measured above the
set PEEP)
Rise Time
Pressure Support
S
(Spontaneous)
Set Mode to SIMV-Pressure or
SIMV-Volume
Set Breath Rate to 0
Set Pressure Support
Inspiratory Time (Apnea backup)
Pressure Control (Apnea backup if
SIMV-Pressure)
Tidal Volume (Apnea backup if SIMV-
Volume)
PEEP
Flow Cycle
Time Cycle
Rise Time
Apnea Rate
CPAP
(Continuous Positive Airway
Pressure)
Set Mode to SIMV-Pressure or
SIMV-Volume
Set Breath Rate to 0
Set PEEP to desired CPAP level
Set Pressure Support to 0
Inspiratory Time (Apnea backup)
Pressure Control (Apnea backup if
SIMV-Pressure)
Tidal Volume (Apnea backup if SIMV-
Volume)
Flow Cycle
Time Cycle
Rise Time
Apnea Rate