Alarms
VentecLife.com 92
The Event Log
The Event Log stores the date and time of unit power on and off, alarm activation, the alarm name, and information
about user interactions with the V*Home alarm system and controls. The most recent 7,300 events, including alarms and
changes to control settings will be accessible through the Event Log. As the V*Home Event Log reaches capacity, the
oldest Event Log records will be overwritten one at a time as new logs are created. Events in the Event Log cannot be
deleted.
NOTE: The Event Log, including information about activated alarms, is retained through V*Home shut down. The Event
Log will be stored for at least 365 days after a total loss of device power.
To view the VOCSN Event Log:
1 Press the Menu tab.
2 Press the LOGS button
3 If needed, press the Events tab.
4 Scroll down on the page to
see additional events. Press
the forward and back page
navigation buttons to see
additional events.
NOTE Timestamps on events in the V*Home Event Log (and Alarm Log) are based on the congured V*Home Date and
Time setting at the time the event occurred. If the V*Home Date and Time settings are incorrect, the timestamps shown
in the log will also be incorrect, and cannot be corrected. However, V*Home Multi‑View timestamps are relative, and will
recalculate based on the most recent V*Home Date and Time setting when data is exported and a report is generated.
NOTE: For more information on V*Home Multi‑View trend reporting,
see “Multi-View” on page 102 .
EXPORT
0 5 10 15 20 25 30L H
No Alarms
High Pressure Alarm Resolved
High Pressure Alarm
Nebulizer Started
Battery Use Alarm Resolved
Battery Use Alarm
Low Breath Rate Alarm Resolved
04/30/2018
04/30/2018
04/30/2018
04/28/2018
04/28/2018
04/25/2018
7:35 PM
7:34 PM
6:05 PM
6:03 PM
6:02 PM
4:13 PM
ALARMSEXIT
EVENTS
PRESSURE (cmH20)
SIMV-VOL
V O
201-250
3674