Limited Warranty
Vermason expressly warrants that for a period
of one (1) year from the date of purchase,
Vermason
Tester Calibration Units
will be free
of defects in material (parts) and workmanship
(labour). Within the warranty period, a unit will
be tested, repaired or replaced at Vermason’s
option, free of charge. Call Customer Service
at
0044 (0) 1462 672005
for a Return Material
Authorisation (RMA) and for proper shipping
instructions and address. Any unit under
warranty should be shipped prepaid to the
Vermason factory. You should include a copy
of your original packing slip, invoice, or other
proof of purchase date. Warranty repairs will
take approximately two weeks.
If your unit is out of warranty, Vermason
will quote repair charges necessary to bring
your unit to factory standards. Call Customer
Service at
0044 (0) 1462 672005
for a Return
Material Authorisation (RMA) and proper
shipping instructions and address.
Warranty Exclusions
THE FOREGOING EXPRESS WARRANTY IS
MADE IN LIEU OF ALL OTHER PRODUCT
WARRANTIES, EXPRESSED AND IMPLIED,
INCLUDING MERCHANTABILITY AND
FITNESS FOR A PARTICULAR PURPOSE
WHICH ARE SPECIFICALLY DISCLAIMED.
The express warranty will not apply to defects
or damage due to accidents, neglect, misuse,
alterations, operator error, or failure to properly
maintain, clean or repair products.
Limit of Liability
In no event will Vermason or any seller be
responsible or liable for any injury, loss or
damage, direct or consequential, arising
out of the use of or the inability to use the
product. Before using, users shall determine
the suitability of the product for their intended
use, and users assume all risk and liability
whatsoever in connection therewith.
UNIT C, 4TH DIMENSION, FOURTH AVENUE, LETCHWORTH, HERTS, SG6 2TD UK
Phone: 0044 (0) 1462 672005, Fax: 0044 (0) 1462 670440 • e-mail: Service@Vermason.co.uk, Internet: Vermason.co.uk
TB-7581 Page 2 of 2
6. Select 10.5M HIGH MID on the
Calibration Unit and repeat the
test. The tester should indicate a
PASS HIGH condition.
7. Select 33M PASS HIGH on the
Calibration Unit and repeat the
test. The tester should indicate a
PASS HIGH condition.
8. Select 37M FAIL HIGH on the
Calibration Unit and repeat the
test. The tester should indicate a
FAIL HIGH condition.
FOOTWEAR CIRCUIT
If applicable
1. Connect the 10mm lead from the
Calibration Unit to the 10mm
socket on the footplate lead of the
High Speed, High Accuracy
Tester using the stud-to-stud
adaptor.
2. Select 715K FAIL LOW on the
Calibration Unit.
3. Press and hold the Calibration
Unit’s test probe on the High
Speed, High Accuracy Tester’s
test plate. The tester should
indicate a FAIL LOW condition.
4. Select 785K PASS LOW on the
Calibration Unit and repeat the
test. The tester should indicate a
PASS LOW condition.
5. Select 9.5M LOW MID on the
Calibration Unit and repeat the
test. The tester should indicate a
PASS LOW condition.
6. Select 10.5M HIGH MID on the
Calibration Unit and repeat the
test. The tester should indicate a
PASS HIGH condition.
7. Select 33M PASS HIGH on the
Calibration Unit and repeat the
test. The tester should indicate a
PASS HIGH condition.
8. Select 37M FAIL HIGH on the
Calibration Unit and repeat the
test. The tester should indicate a
FAIL HIGH condition.
Specifications
Resistance Ranges:
Fail Low 715 kilohm
Pass Low 785 kilohm
Low Mid 9.5 megohm
High Mid 10.5 megohm
Pass High 33 megohm
Fail High 37 megohm
Accuracy:
±2%
Dimensions:
96mm L x 61mm W x 40mm H
Weight:
0.1kg