2
WARNING:
Please read this manual carefully before using the product. Failure to
do so may result in serious injury. SAVE THIS MANUAL
Warning!
Read the operating instructions carefully before use.
Ensure that only trained staff works with the instrument.
Risk of burn!
Caution when touch the housing parts and the hotplate
which can reach temperature of exceed 300 .
Pay attention to the residual heat after switching off.
Protective ground contact
Make sure that socket must be grounded (protective ground
contact) before use.
When operating wear personal safety protection to avoid the risk from:
- Splashing and evaporation of liquids
- Release of toxic or combustible gases
Set up the instrument in a spacious area on a stable, clean, non-slip,
dry and fireproof surface. Do not operate the instrument in explosive
atmospheres, with hazardous substances or around water.
Gradually increase the speed, reduce the speed if:
- Stirring bar breaks away due to high speed
- The instrument is not running smoothly, or the container is not
centered on the plate
Temperature must always be set to at least 50 lower than the flash
point of the media used.
Be aware of hazards due to:
- Flammable materials or media with a low boiling temperature
- Overfilling of media
- Unsafe container
Process pathogenic materials only in closed vessels.