What to do if patient-side temperature data flashes red on my VINCENT Inspired Medical VHB15A Humidifier?
- WWilliam BurnsAug 4, 2025
If the patient-side temperature data display on the LCD flashes red, it indicates that the patient-side temperature is either above the high-temperature alarm limit (invasive 41°C, non-invasive 38°C) or below the low-temperature alarm (invasive 34°C, non-invasive 29°C). If the temperature is too high: * Turn the power off. * Wait for the temperature to decrease. * Turn the power back on. If the red flashing continues, disconnect and use a new unit. If the temperature is too low: * Press the silence alarm to extend the heating time. * Verify that the ambient temperature is not below the operating range. * Check if the air conditioning is making the room too cold.