What to do if VitalGo Total-Lift Bed 425T Medical Equipment bed does not function?
- GGary HendricksAug 14, 2025
If the VitalGo medical bed isn't working, check the power cord connection at both ends and ensure the socket has power. Also, verify that all actuator plugs are securely connected to the controller socket and that the power supply is connected and shows a green light. Check the handset connection to the controller and the cable's condition. In case the bed is tilted or the backrest is raised and CPR is needed, use the manual mode. You may need to clear the error.