Problem Fault
Symptoms:
• Rocking cradle
Possible Causes:
(In probable
order)
• Check for damaged or missing rubber feet.
• If any of the rubber feet are damaged or
missing replace all rubber feet.
Problem Fault
Symptoms:
• Reversed or no volume measurements.
Possible Causes:
(In probable
order)
• Ensure that the flowhead connecting tube is
not pinched or trapped.
Problem Fault
Symptoms:
• Cannot print to external printer.
• Corrupt or missing data on printout.
Possible Causes:
(In probable
order)
• Check that external printer is selected in the
Report Options screen.
• Check USB cable is connected between
In2itive Cradle and printer.
• Check printer as per manufacturers
instructions.
• Check printer compatibility – contact support.
• Main PCB failure – contact support.
Problem Fault
Symptoms:
• Cannot print to PC (Vitalograph Reports
Application).
• Corrupt or missing data on printout.
Possible Causes:
(In probable
order)
• Check that Send to PC is selected in the
Configuration screen.
• Check USB cable is connected between
Vitalograph In2itive and the PC.
• Check to ensure the Vitalograph Reports
Application is correctly installed.
• Check to ensure the required software drivers
are installed on the PC.
• Main PCB failure – contact support.
Problem Fault
• Cannot communicate with Spirotrac V