Vitalograph In2itive 2120 - TECHNICAL SPECIFICATIONS

Vitalograph In2itive 2120
45 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Vitalograph In2itive User Manual
Cat. No. 07514
Issue 5
Page 37 of 45 DT_0006-6
T
ECHNICAL
S
PECIFICATIONS
Product Vitalograph In2itive
Model 2120
Flow detection principle Fleisch type pneumotachograph
Back pressure Less than 0.1kPa/L/second @
14L/s, complies with ATS/ERS 2005
Volume detection Flow integration sampling @ 100Hz
Maximum test duration 90 seconds
Maximum displayed volume 10 L
Volume accuracy
Better than ±3%
Linearity
Better than ±3%
Voltage/Frequency 110-250 V; approximately 50/60 Hz
Accuracy when operated in
operating temperature range
conditions
Flow ±10%
Max. flow rate ±16 L/s
Min. flow rate ±0.02 L/s
Operating temperature range ATS/ERS limits: 17–37ºC
Design limits: 10–40ºC
Performance standards the
Vitalograph In2itive meets or
exceeds
ATS/ERS 2005, ISO 23747:2007 &
ISO 26782 2009
Safety standards EN ISO 60601
QA/GMP standards EN ISO 23747:2007, EN ISO
26782:2009 & FDA 21CFR820
Size 160 mm x 100 mm x 45 mm
Weight 0.23 kg net
Storage Temperature 0–50ºC
Storage Relative Humidity 10%–95%
Printer PCL compatible USB printer.
Communications USB, micro SD Card, Cradle
Notes:
- All values displayed by the Vitalograph In2itive are expressed as
BTPS values.
- Take care not to block the mouthpiece with the tongue or teeth. A
‘spitting’ action or coughing will give false readings.
- Time zero is determined using the back-extrapolated method, from
the steepest part of the curve.