9Safety Information
Vivo 65 users manual
Doc. 006088 En-US U-1b
• All of the physiological alarms of the Vivo 65 must be set at safe lev-
els that will effectively warn the user of any risk. The alarm levels
should be assessed considering the patient settings. Any change of
settings or components may require the readjustment of the alarm
levels.
• Setting the alarm sound level below that of the ambient sound level
can impede recognition of alarm conditions.
• Handle the Vivo 65 with care.
• Do not use the Vivo 65 while in the carry bag.
• Do not use the Vivo 65 with nitric oxide, helium or helium mixtures.
• The Remote Start/Stop accessory is not to be used in Paediatric
mode. Ensure that the Remote Start/Stop is disconnected when
entering Paediatric mode.