37
Error Possible cause Remedy
No operation possible. You are mode After the end of the
„Data transmission“. data tranmissions remove the
SDHC card from the measure-
ment device.
No data memor any more The internal data memory Back up the data on a
available (MEM < 2%). is full. SDHC card and delete the
data memory.
Repairs other than those just described should only be performed by an authorised spe-
cialist.
If you have queries concerning use of the measuring device, our technical support ser-
vice is available at the following telephone number:
Voltcraft®, 92242 Hirschau, Lindenweg 15,
Phone 0180 / 586,582,723 8
Technical data
Operating voltage 230 V/AC 50/60 Hz
Max. power/current 3500 W/15 A
Performance measurement display 0.1 -3500 W
Display energy use 0.000 - 9999 kWh
Display three lines with 4 positions each
Tariff range 0,000 - 9,999
Accuracy 5 - 3500 W (± 1% + 1 count)
2 -5 W (± 5% + 1 count)
< 2 W (±15% + 1 count)
Buffer battery 3 V, CR1620
Ambient conditions 10 - 50 °C/max. 90%rH (not condensing)
Operating altitude: max. 2000 m (above MSL)
Weight ca. 240 g
Dimensions (LxWxH) 164 x 82 x 83 (mm)
Overvoltage category CAT II
Pollution degree 2
Measurement tolerances
Statement of accuracy in ± (% of reading + display error in counts (= number of smallest points)). The accu-
racy is valid at a temperature of +23°C ± 5°C, and at a relative humidity of less than 75 %, non-condensing.
In no event exceed the max. permitted input values.