VOLTECH AT3600 USER MANUAL CREATING THE PROGRAM
TEST PROGRAM EDITOR VPN 98 - 024 PAGE 3.3.29.
Hex Code
Error
0000
Test passed with no errors.
xxx1 Voltage over-range occurred.
xxx2 Current over-range occurred.
xxx4 Ramp-up process has been aborted; too much load.
xx2x
The test parameters could not be trimmed to required value; incorrect
test parameters for the load.
xx4x
An error occurred after ramp-down; current did not decay away in
time.
x1xx
Current limit fail (hi-pot tests only); test current exceeded pre-
programmed limit (AT3600 only).
x2xx
Voltage present (Vpres) did not go away in time; something on the
fixture is staying charged (AT3600 only).
#4xx
The test has failed due to an interrupt (see table below).
Where # = reason for interrupt fail.
1
The safety interlock interrupt has terminated the test; a break has
occurred from the safety interlock port (AT3600 only).
3
A hi-pot trip interrupt has terminated the test; a flashover has occurred
or the current has exceeded the AT’s current capabilities (can occur on
HPAC, HPDC, ACRT, DCRT and IR) (AT3600 only).
4
A STOP interrupt has terminated the test from the STOP flag of the
remote port.
5
An 8kV interrupt has terminated the test - voltage > 8kV (AT3600
only).
7
An over-temperature interrupt has terminated the test; unit has become
too hot (AT3600 only).
8
An auxiliary trip interrupt has terminated the test; this can be created
by an attached device to the peripherals port.