EasyManua.ls Logo

Voltech AT3600 - Page 127

Default Icon
424 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
VOLTECH AT3600 USER MANUAL PROGRAMMING TESTS
TEST PROGRAM EDITOR VPN 98 - 024 PAGE 3.4.21.
What happens if I set both absolute limits for both primary and secondary
windings?
If you want to detect up to a 10 turn error on the primary and 1 turn on the secondary
in the above example, set 91/2 (pri) and 1/2 (sec).The tester uses the equivalent ratio
limits: 21.95 and 18.05. Setting both limits WIDENS the limit to allow for an
error of 10 Turn on the primary SIMULTANEOUSLY with 1 turn on the
secondary.
i.e. to allow 190:11 (17.27) and 210:9 (23.33).
Notice that these limits will NOT detect just a 1 turn error on the secondary because
you have now chosen limits to allow for errors on both primary and secondary,
resulting in a broader tolerance on the absolute ratio.
How so I set limits that will detect a 1 turn error in either the primary or
secondary winding?
To do this you must set limits that will detect the smallest possible error. This will be
due to the 1 turn error being on the winding with the largest number of turns.
In the above example then, set absolute limits of ½ turn (pri) and 0 turn (sec). (
Remember to use the n-1/2 rule).
The equivalent limits used by the tester will be:
20.05 and 19.95
This will certainly detect a 1 turn error in the primary, but because of the physical
errors already described these limits may be too severe for a normal good part. This is
why a combination of inductance and turns ratio tests are normally used to confirm
the correct assembly of transformers.
The polarity test may be selected as (+)ve, (-)ve or no test. To measure the exact
interwinding phase angle use the PHAS test.
Use the PC Editor’s ‘measure’ button (together with a sample part) if you are unsure
of the nominal value.

Related product manuals