VOLTECH AT3600 USER MANUAL PROGRAMMING TESTS
TEST PROGRAM EDITOR VPN 98 - 024 PAGE 3.4.23.
3. Select the high and low terminals of the primary winding. (Click on the arrow to
the right of the fill-in box and you will be provided with a choice of the available
terminals. Click the required terminal.)
Select the high and low terminals of the secondary winding.
4. Select (mouse click the button) Ratio (%), or Turns (#) limits.
Enter the required limits:
Ratio (%)
Enter the nominal Primary to Secondary ratio, together with the limits as
positive and negative percentage values.
Turns(#)
In turns mode you may set upper and lower limits on the primary and
secondary windings. The transformer tester will then set limits on
the turns ratio which are equal to the sum of these two errors, e.g.:
Primary 100 turns ± 2 turns
secondary 100 turns ± 2 turns
the tester will set the following limits on the ratio:
upper limit = 104/100 lower limit = 96/100.
If you wish to set limits on only one winding, you must enter zero (0) as the
limit for the other winding.
Note that the Measure button may be used here as outlined in section 3.9.
5. If required, the polarity of the test can be checked. Click on the expected polarity
of the transformer turns, otherwise disable the test by selecting ‘No Test’.
6. Select OK. The test specification will be displayed in the program window on
the screen.