PROGRAMMING TESTS VOLTECH AT3600 USER MANUAL
PAGE 3.4.60. VPN 98 - 024 TEST PROGRAM EDITOR
7. Choosing the test limits.
The test limits can be:
% A nominal value together with + and – percentage limits.
>< Minimum and maximum values
> A minimum value only
< A maximum value only
Use the Editor’s ‘measure’ button (together with a sample part) if you are
unsure of the nominal value.
Remember to click on the units of limits (mV or V) when entering absolute
values as limits.
Test limits should be set as close as possible to the nominal value in order to
detect faults. A typical tolerance would be ± 5%