PROGRAMMING TESTS VOLTECH AT3600 USER MANUAL
PAGE 3.4.66. VPN 98 - 024 TEST PROGRAM EDITOR
To program the test:
1. Enter the signal, frequency and bias current required for the test.
Note that the signal is normally entered as a voltage by selecting the ‘mV’ or ‘V’
button; but you may optionally enter it as a current by selecting the ‘mA’ button.
(If in doubt refer to the ‘Test Conditions’ chapter of your AT user manual.
Note also that the measure button may be used here, as outlined in your AT user
manual, with the combinations of test parameter specification shown in the
following table:
Option 1 Option 2 Option 3 Option 4
Signal
Specified
voltage
Specified
current
Auto Auto
Frequency
Specified Specified Specified Auto
Bias
Specified Specified Specified Specified
2. Select (mouse click the button) the integration time you require.
‘Medium’ is the default setting. ‘Long’ will give you more stable readings (for
tighter limits) at the expense of test time, and ‘Short’ will test at the maximum
speed, but may give slightly noisier readings.
3. Enter the terminal names between which the inductance will be measured.
By pulling down the menu along side the fill-in box, you may select the high and
low terminals for each test.
4. Select (mouse click the button) the type of limits you require, and enter the
values.
% Enter a nominal value and then the limits as negative and positive values.
>< Enter minimum and maximum values.
> Enter just a minimum value.
< Enter just a maximum value.
5. Select OK. The test specification will then be displayed in the program
window.