FIXTURE COMPENSATION VOLTECH AT3600 USER MANUAL
PAGE 6.4.2. VPN 98 - 024 FRONT PANEL OPERATION
OPERATOR INSTRUCTIONS
After you have entered the name of a program where compensation can be applied, at
the next display:
• Fit the shorting bobbin into the test fixture on the tester.
• Press the COMPENSATE soft-key.
The tester will perform the short and open circuit compensation.
As the compensation measurements are being made, the display will briefly show:
The tester will then show:
COMPENSATION PASS AND FAIL INDICATIONS
In the event of a failure check the following:
• The short circuit connections (if applicable)
• The fixture integrity.
And then retry the compensation.
If desired the test programs can be run without fixture compensation being
performed, however this may affect the accuracy of the results, depending upon the
type of transformer being tested and the fixture design.