TRANSFORMER TESTS VOLTECH AT3600 USER MANUAL
PAGE 7.1.6. VPN 98 - 024 TESTS AND TEST CONDITIONS
TEST CURRENT
The test current is set according to the value specified as maximum in the test limits:-
Maximum
Resistance
<0.1Ω <1Ω <10Ω <100Ω <1kΩ <10kΩ
Test current
400mA 200mA 100mA 30mA 3mA
300µA
NOTE: Above 10kΩ, a constant voltage source of 8.0V is used.