VOLTECH AT3600 USER MANUAL SPECIFICATION
SPECIFICATION VPN 98 - 024 PAGE 8.1.15.
Special Notes:
Under certain test conditions the measured current will be higher than the allowed
minimum limit due to parasitic effects on the AT3600 and any connected fixture system.
The maximum parasitic effect created by the AT3600 with every node being used on an
HPAC test can be calculated by:
I
p
= V*(4.614*10
-10
+ 9.1*10
-11
*f) where V = Test voltage and f = test frequency.
This works out to be 12.5uA at 2.5KV @ 50Hz.