CREATING THE PROGRAM VOLTECH AT3600 USER MANUAL
PAGE 3.3.14. VPN 98 - 024 TEST PROGRAM EDITOR
3.8. TEST LIMITS
In the Quick Start Tutorial (section 3 of chapter 2), it was stated that for the R -
Winding Resistance - test the limits could be entered in four different ways. The R
test is not unique in this respect. In fact, most tests have several ways in which the
limits may be entered, allowing you greater flexibility in configuring each test to
match your production testing requirements.
The four types of limits for the R- Winding Resistance - test are:
%: The limits are entered as positive and negative percentages about a nominal
value.
> <: The limits are entered as minimum and maximum values.
> : There is only a minimum limit.
< : There is only a maximum limit.
The corresponding dialogue boxes are as follows: