CREATING THE PROGRAM VOLTECH AT3600 USER MANUAL
PAGE 3.3.22. VPN 98 - 024 TEST PROGRAM EDITOR
RESTRICTIONS FOR AUTOMATICALLY OBTAINING THE TEST PARAMETERS
As shown above, the LS - Winding Inductance (Series Circuit) - test is one where
both the test voltage (or current) and frequency may be obtained automatically by
using the Measure button.
Alternatively, you may wish to specify yourself either one or both of the parameters
before using the Measure button. In this case, the tester will attempt to make use of
the parameter(s) that you have specified, and enter a value only for parameters that
you have not specified.
In general, all tests have restrictions on which test parameters may be (or have to be)
specified, and which may be chosen automatically. In the case of the LS - Winding
Inductance (Series Circuit) - test, the valid combinations are summarized in the
following table:
Signal
Specified
voltage
Specified
current
Auto Auto
Frequency
Specified Specified Specified Auto
(The combination that is not allowed, is to specify the voltage (or current), and ask
the AT to find the frequency.)
If you have entered a non-permitted combination, the Editor will display a warning
message when you click on the Measure button, and there will be no measurement
until you correct the error.