Guidance and Manufacturer’s Declaration – Electromagnetic Immunity (continued)
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environmental -
Guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 V
3 V/m
Portable and mobile RF communications
equipment should be used no closer to any
part of the Wallach Colposcope, including
cables , than the recommended separation
distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
P
V
d
1
5.3
P
E
d
1
5.3
80 MHz to 800 MHz
P
E
d
1
7
800 MHz to 2.5 GHz
where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in meters
(m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site
survey,
a
should be less than the compliance
level in each frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following symbol:
a
b
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the Wallach Colposcopes
are used exceeds the applicable RF compliance level above, the Wallach Colposcope should
be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating the Wallach Colposcope.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
IMSC036-DFU • Rev. B • 10/14 18 Made in the USA